Caution:
When lifting the flow flange from the reactor, three technicians should
be utilized. One technician runs the lift and two spotters watch and stabilize the flow flange
to ensure that it lifts evenly without bouncing.
14. Carefully pull the lift assembly away from the reactor chamber.
15. Using the hydraulic lift assembly, raise the flow flange assembly high enough for the shutter to safely
clear the top of the reactor walls.
Warning!
To avoid having the flow flange strike the top of the system frame, do
not raise the flow flange too high.
16. Carefully pull the lift assembly away from the reactor chamber.
Figure 5-29. Manual Isolation Valves and Connections