Figure 5-128. Particle Filter Neck Collar (Green Highlight)
12. (This step requires two people; on either side of the particle filter housing.) Stand the particle filter
housing on its feet until it is perpendicular to the floor and able to stand alone.
Note:
In proceeding with particle filter maintenance you may opt to use the oxidation valve
at the bottom of the filter housing. If deciding against this option, skip to Step 14 of this
procedure. To utilize the option, connect the N2/Air source at a 0.97/0.03 ratio, at 5 PSI, to
the valve. Open the oxidation valve slowly. Once fully open, open the filter outlet ball valve
slowly
to a vented housing that will cover the KF40 fitting
. Allow
slow
oxidation for
one hour.
Warning! Watch for fire or flare ups.
In the even of a fire or flare up, place
the reducer back on the body to starve the reaction of oxygen. Repeat this process until all
reactions have ceased.