5. Remove the confined inlet from the flow flange weldment. This component is very heavy. Two or more
people should complete this task.
6. Replace two of the screws at 180 degrees from each other.
Note:
These screws will not tighten fully, but they will hold the cold plate in place when the
Flow Flange is rotated 180 degrees.
7. Rotate Flow Flange 180 degrees, so that it is
right-side-up
.
8. Lower the Flow Flange onto two parallel supports with the two screws between the supports, and the
Flow Flange supported on the supports.
9. Remove the two screws.
10. Raise the Flow Flange to remove the cold plate.
11. To remove the gas screen from the coldplate, loosen and remove the vented screws securing the
screen.
12. Remove the screen from the coldplate.
13. Rotate the Flow Flange assembly 180 degrees.
14. Remove the baffled section.