K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-66
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
Figure 5-61. Preparing O-Rings
10. Once all the o-rings have been wiped clean, place a small amount of Fomblin on one palm, take a
small swab then apply the Fomblin on the o-rings. Take care not to apply too much Fomblin, leaving
only a very thin film.
Figure 5-62. Fomblin on O-Rings
11. Take note of the seams on the o-rings, and make sure to position them to run along the sides during
the o-ring installation. The seams must not be in a position to run along the top nor bottom surfaces,
as this would increase the risk of leakage. Install o-ring in the inside groove and install o-ring in the
outside groove.