K465i Operations and Maintenance
P/N 1212831 Rev C
Page 1-12
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
Robot Pushbutton
The Robot mechanism is used to transfer a platter
containing wafers, into and out of the reactor chamber
and glovebox.
During initial setup the robot must be
taught
positioning and the pushbutton is used to immediately
stop the robot moving in a situation such as over-
travel. It also can be used when determining that the
robot mechanism is not functioning correctly during
operation.
To enable robot operation the activated Robot
pushbutton must be pulled out and then operation is
enabled from the control computer.
Alarm Buttons
The
Alarm Silence
button disables a
Hardware
Audible Alarm
trigger. The
Alarm Reset
button
resets the hardware safety relays.
Table 1-3. Hardware Alarm Functions
ALARM
SILENCE
ALARM
RESET
ALARM STATUS
Lights OFF
OFF
OK
Lights ON
OFF
Hardware or Software
1.
Clear alarm in Nexus
If alarm remains on, it is a hardware alarm.
2.
Push the Alarm Silence button.
Lights ON
ON
Hardware Alarm Reset Required
1.
Push Alarm Reset
If lights remain on, the Hardware Alarm Input is
still on.
2.
Check Nexus, from the Alarm that is LOW.