K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-170
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
Section 5.10.
Gas/Hydrogen Monitoring
Veeco's TurboDisc systems are equipped with a hydrogen detection apparatus which warns
the operator of a hydrogen leak.
Safety Precautions and Hazard Management
In addition to the technician responsible for performing the maintenance, a second
site-qualified technician must be present in the FAB during the maintenance
procedure.
Whenever maintenance or other system related functions are performed on Veeco
TurboDisc systems, safety procedures must be followed, including the wearing of all
required personnel protection equipment.
A 4% or higher concentration of hydrogen in the air is flammable. The test or 'span'
gas typically used to calibrate the hydrogen detector is 50% LEL or 2% hydrogen.
Cleanroom compatible gloves should be worn when performing this maintenance.