Placement of the Middle Filament
30. Supporting the middle filament from both sides of its center with both hands, gently lower the filament
onto the wire supports, lining up the connection ends with the electrodes. Ensure that the filament is
placed with the coated (darker) side facing up.
31. Place clamps over the connector ends and install the tungsten screws finger tight. Maintain the
alignment of the threaded clamp plate under the flexible electrode by stabilizing it with one hand,
making it possible to insert the screws properly during assembly.
32. Using a caliper as a measurement device, adjust the filament from left to right until it is equidistant
from the sides of the water-cooled baseplate.
Note:
The filament is not concentric with the baseplate in the front-to-back direction,
looking from the electrodes’ end. Ensure that the edges of the filament connector ends are
flush with the edges of the electrodes.
33. Using the torque screwdriver and the flat head bit, gradually tighten all the screws for the Middle
filament until a torque value of
10 in-lbs
is reached for each screw. First tighten the screws on the
fixed electrode; hold and prevent the flexible electrode from twisting with one hand, while tightening
the other two (2) screws with the other hand (to the specified torque).
Installing the Outer Filament
34. In the same manner as for the Middle Filament, unpack, install (see Figure 5-114), align and position
the Outer Filament relative to insulation blocks (see Figure 5-114). Then using the torque screwdriver
with the flat head bit, gradually tighten the screws of the Outer filament (first to the non-flexible
electrode and then to flexible electrode, preventing it from twisting) until a torque value of
10 in-lbs
is reached for each screw.