K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-98
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
Section 5.5.
K465i GaN Heater Maintenance
This section covers the removal and replacement of the K465i GaN heater assembly. All
steps preceding the start of heater maintenance are covered in other procedures and are
not covered in this section. Please refer to appropriate sections for the applicable
procedures.
Note:
Detailed heater assembly/disassembly
will be performed by Veeco Instruments
personnel only
. This document includes filament and filament-supports
removal/replacement and whole heater removal/replacement.
Safety Precautions and Hazard Management
In addition to the technician responsible for performing the maintenance, a second
site-qualified technician must be present in the FAB during the maintenance
procedure.
Whenever maintenance or other system related functions are performed on Veeco
TurboDisc systems, safety procedures must be followed, including the wearing of all
required personnel protection equipment.
An electrical shock hazard exists. De-energize heater power supplies before servicing
heater assembly.
The spindle assembly is a rotation hazard. De-energize the spindle motor before
servicing heater assembly.
Cleanroom compatible gloves should be worn when performing this maintenance.
Disposable clean room sleeves should be utilized to protect arms/clothing from
contamination.
PROCEDURES IN THIS SECTION
Tools and Materials
Heater Disassembly Guide
Heater Cleaning Procedures
Heater Reassembly Guide