K465i Operations and Maintenance
P/N 1212831 Rev C
Page 1-8
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
SAFETY INTERLOCKS AND ALARMS
The following safety interlocks and alarms (
) are present in the K465i GaN. Most
system sensors generate a software alarm condition that shuts the system down. For some
critical systems there is a hardware shutdown (EMO), independent of the software shutdown
that is initiated upon an alarm condition.
Chapter 5: Maintenance and Repair
of this
document includes information regarding the calibration and testing of these safety alarms.
Table 1-2.
Safety Interlocks and Alarms
SYSTEM COMPONENT
ALARM SET POINT
ACTION ON ALARM
H2/N2 Line Pressure Low
10-12 psi
EpiView or Nexus System
Shut Down
Alkyl Manifold Pressure Low
10-12 psi
EpiView or Nexus System
Shut Down
Pneumatic Pressure Low
70 psi
EpiView or Nexus System
Shut Down
Process Pump Exhaust Over
Pressure
2 psi
EpiView or Nexus System
Shut Down
Cabinet Ventilation Pressure
Low
0.5" water
EpiView or Nexus System
Shut Down and
Activate Gas Panel Minimizer
Bank
H2 Detector
0.5 LEL
EpiView or Nexus System
Shut Down and
Activate Gas Panel Minimizer
Bank
Hydride Gas Leak
System-
Dependent
EpiView or Nexus System
Shut Down and
Activate Gas Panel Minimizer
Bank
H2/N2 Line Pressure Low
10-12 psi
EpiView or Nexus System
Shut Down
Alkyl Manifold Pressure Low
10-12 psi
EpiView or Nexus System
Shut Down
Pneumatic Pressure Low
70 psi
EpiView or Nexus System
Shut Down
THA Over Pressure
40
torr>atmospheric
60
torr>atmospheric
EpiView or Nexus System
Shut Down
Hard Shut Down/de-energize
the PLC
Growth Chamber Over
Pressure
40
torr>atmospheric
60
torr>atmospheric
EpiView or Nexus System
Shut Down
Hard Shut Down/de-energize
the PLC