K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-44
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
19. Using Cleanroom wipes, gently wipe the bottom of the flow flange - with the sole purpose of removing
loose dust only (NO IPA).
Figure 5-40. Wiping of Flow Flange Bottom
20. Lower the shutter and wipe the chamber walls using dry clean room wipes only (No IPA).
21. Raise the shutter.
22. Vacuum and then slowly wipe the liner for the heater assembly using dry Cleanroom wipes(No IPA).
23. Reinstall the liner for heater assembly ensuring that it is seated correctly in the groove.