System Description
ARCADIS Varic
SPR2-310.620.30.02.02
35
of 44
Switching the system off: complete shut down in the main
menu
0
◆
Select
Options > End Software Session
in the main menu at the monitor
trolley.
– The
System Message
dialog box is displayed.
◆
Click
Shutdown
.
– The imaging system is completely shut down.
◆
Press the
OFF
switch on the monitor trolley.
– The C-arm system, the imaging system and the monitors are switched off.
Using the
End Software Session
menu command (or automatically upon
16th switch off using the OFF button), the imaging system is completely shut
down before it switches off. When the ARCADIS Varic is switched on again after
a complete shutdown, it may take more than 3 minutes, depending on the con-
figuration and previous use, for it to reach operational readiness.
The
Restart
button in the
System Message
dialog box only shuts down the syn-
go user interface and the application programs and then restarts them. This can
be used, for example, when installing a virus protection update (
→
Register 10:
Maintenance, Page 10).
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...