Operator Manual
24
of 52
SPR2-310.620.30.02.02
Examination
Using the fill image
from subtraction
angiography
Prerequisites:
A subtraction angiography (SUB operating mode) was performed at any time dur-
ing the current examination.
The symbol for an existing subtraction mask is shown in the control area.
◆
Press the
RoadMap
key on the control panel
once
.
– The operating mode is switched to
Roadmap
.
– If the vessels are easy to recognize, you can now start positioning the cath-
eter.
If you press the
RoadMap
key twice, the existing subtraction image is discarded.
You will then have to regenerate the fill image.
Positioning the
catheter
◆
Release radiation with the hand switch or with the assigned footswitch.
– The live image is displayed on the left monitor.
◆
Position the guide wire or the catheter under fluoroscopic control.
– The right monitor shows the fluoroscopic image.
– The left monitor shows the subtracted Roadmap image with catheter.
You can repeat fluoroscopy as often as needed while you insert the catheter.
If you select
RoadMap
again, you will start a new roadmap scene.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...