Options
ARCADIS Varic
SPR2-310.620.30.02.02
23
of 24
Deleting the diaphragm positions
0
◆
If you wish to delete the stored diaphragm positions again, press one of the
diaphragm keys and the diaphragm will open completely when the cassette is
inserted.
Setting exposure data
0
If a cassette is loaded after the system is restarted, only dashes will be displayed
in the kV and mAs fields. You cannot release an exposure since this can only be
done after setting the required kV and mAs values.
◆
Now press one of the kV or mAs +/- buttons.
– The preference values 40 kV and 5 mAs will be displayed. You can then
change these values.
◆
Now press one of the kV or mAs +/- buttons until the required values are set.
– If you perform fluoroscopy in the meantime, the kV and mAs values used
for the last cassette exposure will remain set when reloading the cassette.
Example of kV and mAs values:
Display of kV values
Display of mAs values
kV
mAs
mA
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...