Operator Manual
4
of 34
SPR2-310.620.30.02.02
Technical Data
SIREMATIC curves
0
kV/mA curves for Fluoroscopy and Pulsed Fluoroscopy
(0.5 to 2 frames/s)
0
(1) LD 330 W
SIREMATIC LD
Low Dose
standard mode max. 3.0 mA
in Power Mode 600 W max. 5.5 mA
(2) S1 330 W
SIREMATIC S
Standard
standard mode max. 3.0 mA
in Power Mode 600 W max. 5.5 mA
(3) S2 550 W
SIREMATIC S
Standard
standard mode max. 5.0 mA
in Power Mode 1000 W max. 9.1 mA
(4) HC1 330 W
SIREMATIC HC
High Contrast
standard mode max. 4.7 mA
in Power Mode 600 W max. 8.5 mA
(5) HC2 550 W
SIREMATIC HC
High Contrast
standard mode max. 7.8 mA
in Power Mode 1000 W max. 14.3 mA