System Description
ARCADIS Varic
SPR2-310.620.30.02.02
23
of 44
Selecting the operating mode
0
You can select the following operating modes for the ARCADIS Varic:
Fluoroscopy
For fluoroscopy (CFC) you can choose between several exam sets with different
characteristic curves to determine exposure parameters for fluoroscopy. Every
application allows to choose between different optimized programs. The stan-
dard setting of the ARCADIS Varic after startup is the basic setting configured for
the body region "All", which is typically
Continuous fluoroscopy
(
→
Register 8:
Configuration, Page 5).
Exposure factors and system control units including the way in which the auto-
matic setting is controlled:
❏
1 K
2
matrix; 30 f/s frame rate; image integration (as a function of the K factor
set), i.e. a number of K exposures are integrated into one image by sliding
averaging; the K factor can be selected between K = 1 ("OFF" setting) and
K = 32 and can be assigned to an exam set and stored.
Typical clinical procedure:
❏
Fracture reposition of the distal upper extremity (e.g. distal forearm fracture)
in the plaster room of an emergency outpatient clinic where, under continuous
fluoroscopy, the fracture elements are reduced by extension, fixed temporari-
ly in the best possible position and then fixed permanently by applying a plas-
ter cast.
Pulsed fluoroscopy
This operating mode (frame rates up to 15 f/s) allows a reduction in the radiation
dose of up to 70% for the patient and operator. The pulse duration is generally
7 milliseconds. According to the level of noise reduction, many different fluoro-
scopic images can be integrated. For frame rates less than or equaling 2 frames
per second, a type of intermittent continuous fluoroscopy is used where the
pulse duration varies depending on the noise reduction set.
Exposure factors and system control units including the way in which the auto-
matic setting is controlled:
❏
1 K
2
matrix; typical frame rate 4-15 f/s; image integration (as a function of the
K factor set), i.e. a number of K exposures are integrated into one image by
sliding averaging; the K factor can be selected between K = 1 ("OFF" setting)
and K = 8 and can be assigned to an exam set and stored.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...