Operator Manual
32
of 44
SPR2-310.620.30.02.02
System Description
Confirming a warning
signal
After 5 minutes of fluoroscopy time, a warning signal sounds. This can be reset
at the control panel. If no reset is performed, radiation is automatically disabled
after another 5 minutes (total of 10 minutes after the last reset).
These settings are country-specific and can be changed by Siemens Service in
accordance with the applicable regulations.
◆
Press the
Reset key
on the control panel of the C-arm system.
– The acoustic warning signal is deactivated.
Storing images
(during radiation)
◆
Press this button on the hand switch during radiation.
– The image currently generated and displayed is saved.
(
→
Register 5: Examination, Page 17)
Storing images
(after radiation)
◆
Press this button on the hand switch.
– Holding the key for < 2 seconds: saves the image last recorded (LIH).
– Holding the key for > 2 seconds: saves the scene last recorded (LSH).
The ARCADIS Varic transfers images from monitor A to monitor B and then
stores them in the local database.
Images/scenes can also be stored via the corresponding button on the optional
multifunctional footswitch, if used.
(
→
Register 11: Options, Page 8)
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...