Operator Manual
14
of 28
SPR2-310.620.30.02.02
Safety
Personal safety
0
Open heart and skull examinations
0
If an approved system is used alone or with other equipment for open heart or
open skull examinations, a conductive connection must be made between the
system and a potential equalization point, e.g. the tabletop
(
→
Register 3: System Description, Page 10).
Only then can the patient be connected to the system.
Crushing hazards on the C-arm system
0
Correct handling of the C-arm system requires that operating personnel and
patients use only the grips provided for this purpose. Where this is not possible,
monitor the points of potential crush injury between movable system parts and
their guide openings.
Warning
Moving and braking the C-arm (see Fig. 1 and 3).
Risk of crushing hands!
◆
Please make sure that your hands are not in the travel path of system parts.
Warning
Maximum lowering of the C-arm (see Fig. 2).
Risk of crushing feet!
◆
Please watch your feet when the C-arm is being lowered fully, since there
may not be sufficient clearance left between the I.I. and the floor.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...