Operator Manual
30
of 44
SPR2-310.620.30.02.02
System Description
Radiation release
0
Radiation can be released with the hand switch or the footswitch.
Complete image integration as a function of the set k factor (up to max. k = 8,
depending on the operating mode) is ensured even for very short exposures
("toe tapping").
k factor: a number of k exposures are integrated into one image; the k factor can
be selected between k = 2 and k = 32 and can be assigned to an exam set and
stored (IEC 60601-2-7:1998 29.1.103 d).
Key switch
0
To ensure that radiation is released by authorized personnel only, the C-arm in-
cludes a key switch.
The key switch is located on the left side of the C-arm (viewed from the direction
of travel).
◆
Prior to releasing radiation, turn the key switch to the right until it is horizontal.
– Radiation now can be released either with the hand switch or the foot-
switch.
◆
Return the key switch back to its original position.
– Radiation can no longer be released with either the hand switch or foot-
switch.
– The key can now be removed.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...