System Description
ARCADIS Varic
SPR2-310.620.30.02.02
33
of 44
Radiation release with the footswitch
0
The footswitch is used if both hands need to be free during the exposure.
Standard footswitch
The footswitch is also suitable for applications where fluids may land on the floor.
For operation of the multifunctional footswitch (option) see
(
→
Register 11: Options, Page 8).
Releasing radiation
The right footswitch is always used to activate fluoroscopy (CFC) (standard set-
ting).
The left footswitch is used to activate the currently selected operating mode.
Exception
: If fluoroscopy (CFC) is selected, the left pedal is assigned the digital
radiography (DR) mode.
The functionality of the two pedals can be reversed by Siemens Service upon
request.
◆
Keep the foot pedal pressed during radiation release.
SIEMENS
SIEMENS
SIEMENS
SIEMENS
Warning
When the C-arm is rotated by 180° and fully lowered, the I.I. and the footswitch
may come into contact.
Unintentional release of radiation!
◆
Please make sure that the footswitch is not located underneath the I.I.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...