Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02
9
of 34
Dosimetric information
0
3rd edition IEC 60601-2-54
Air kerma strength
0
The skin dose values were measured at a distance of 30 cm from the image
intensifier input with a 20 cm PMMA phantom pursuant to IEC 60601-2-54:2009,
203.5.2.4.5.101.
*) standard examination program provided
Program*
Curve
Format
Dose
Air kerma reference
CFC HC2
Normal
medium
19
mGy/min
Standard
medium
push
20
mGy/min
CFC IOD
Normal
low
12
mGy/min
Standard
medium
18
mGy/min
high
22
mGy/min
Card
IOD
Normal
high push
33
mGy/min
Program
*
Curve
Format
Noise
reduction
Air kerma reference
DR
DR
1000
Normal
k = 4
0.2
mGy/f (image)
k = 8
0.4
mGy/f (image)
k = 16
0.8
mGy/f (image)
k = 32
1.6
mGy/f (image)
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...