Operator Manual
18
of 52
SPR2-310.620.30.02.02
Examination
Storing images
(after radiation)
◆
Press this key on the control panel of the C-arm system.
– or –
◆
Press this key on the hand switch.
– Holding the key for < 2 seconds: saves the image last recorded (LIH).
– Holding the key for > 2 seconds: saves the scene last recorded (LSH).
– The saved LIH is displayed on the right (reference) monitor (
→
Page 25).
Changing the image
display
The control panel of the C-arm system provides basic tools for optimizing the
image display of the
Examination
task card.
◆
Edge enhancement (4 levels).
◆
Selection of LUT steps on monitor A.
◆
Enlarging/reducing the image.
During an examination, single frames and scenes can automatically be loaded
into the
References
card (
→
Page 31). Changes in edge enhancement only affect
the left monitor (the image is displayed with an enhanced or less enhanced edge).
The reference image on the right monitor is not affected. If the difference
between the images on the left and right monitor is a distraction, it is recom-
mended not to change the edge enhancement at the control panel of the C-arm
system during the examination.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...