System Description
Connecting the footswitch
0
For the release of radiation with the footswitch, it must be connected to the
C-arm system.
◆
Plug the footswitch cable into the socket labeled with the footswitch symbol
at the C-arm system.
◆
Open the flap for cabling.
◆
Place the footswitch cable inside the groove for the cable and close the flap.
Establishing the equipotential bonding connection
0
The ARCADIS Varic can be connected to a protective ground terminal via the
equipotential connector on the C-arm system. This will ensure that the
ARCADIS Varic has the same electrical potential as other units connected to the
same protective earth terminal.
When performing cardiac examinations or examinations of the open skull, an addi-
tional grounding cable according to DIN 57107/VDE107 must be routed in rooms of
Application Group 2.
◆
Clamp the grounding cable to the front face of the C-arm system (arrow) and
to an equipotential bonding point in the patient vicinity.
– Equipotential bonding is established.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...