Operator Manual
26
of 44
SPR2-310.620.30.02.02
System Description
Continuous
Fluoroscopy (CFC)
◆
Press this key.
–
Fluoroscopy
mode is selected.
– The LED lights up.
Upon startup of the ARCADIS Varic, the basic setting configured for the body
region "All", which is usually
Fluoroscopy
, is used as the automatic default
(
→
Register 8: Configuration, Page 5).
Pulsed
Fluoroscopy (PFC)
◆
Press this key.
–
Pulsed Fluoroscopy
mode is selected.
– The LED lights up.
Digital
Radiography (DR)
Digital Radiography (DR) provides an electronic instant image with best image
quality. DR is recommended for final exposures.
On activation of digital radiography a short radiation pulse is released.
◆
Press this key.
–
Digital Radiography
mode is selected.
– The LED lights up.
In order to ensure sufficient image quality, images are completely acquired after
the start of acquisition, even if no more radiation is released. The effective acqui-
sition time depends on the preset integration factor and is a maximum of 1.8 sec-
onds.
After complete image acquisition, radiation is automatically switched off, even if
the radiation release button remains pressed.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...