Operator Manual
10
of 34
SPR2-310.620.30.02.02
Technical Data
Maximum air kerma strength
0
The skin dose values were measured at maximum generator power at a distance
of 30 cm from the image intensifier input pursuant to IEC 60601-2-54:2009,
203.5.2.4.5.101.
*) standard examination program provided
**) In ADR stop mode, the maximum kV value is set
Program*
Curve
ADR stop**
Air kerma reference
CFC HC2
110 kV/5 mA
52
mGy/min
Standard
HC1
110 kV/3 mA
32
mGy/min
Program*
kV
mAs
Air kerma reference
Cassette
110
69
12
mGy/f (image)
Exposure
72
mGy/min