Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02
17
of 34
Current/voltage values
Generation of radiation
0
Generator
0
Voltage
Long-term current consumption
Short-term current consumption
100 V
110 V
120 V
127 V
200 V
230 V
240 V
17 A
16 A
15 A
14 A
9 A
8 A
8 A
29 A
26 A
24 A
23 A
15 A
13 A
12 A
Pulsed Fluoroscopy
Min. pulse width 7 ms
Pulse rate 15 f/s/23 mA
Max. pulsed output 2.3 kW
Fluoroscopy
40 kV to 110 kV/0.2 to 15.2 mA (max. 1000 W)
Digital Radiography
40 kV to 110 kV/0.2 to 23 mA (max. 1000 W)
Current-time-reference
product
1.6 mAs/102 kV
Tolerances
kV
±
10% (measured with spectrometric kV method)
mA
±
8%
±
0.1 mA (measured in rectified high-voltage circuit)
Fluoroscopic time 1 digit (6 s)
±
5%
Power rating
1.4 kW (102 kV/1.6 mAs)
Min. pulse width 7 ms
Max. pulsed output 2.3 kW
Exposure times
min. 50 ms at 40 kV 20 mA
max. 10 s
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...