Shutdown
0
Before disconnecting the ARCADIS Varic from the mains, you must shut it down.
During the shutdown procedure the imaging system is powered down. There are
two different procedures, each with a different reset time.
The shutdown procedure used depends on how you switch off the
ARCADIS Varic:
Simple shutdown
0
◆
Press the
OFF
switch on the monitor trolley.
– The C-arm system is immediately switched off.
– The imaging system and monitor A switch off after the system has been
shut down.
The image system is shut down to "hibernate" before it is switched off by using
the
OFF
button. When the ARCADIS Varic is switched on again, it will normally
be ready for operation after approx. one minute.
Depending on the configuration, it may take longer than one minute for the
ARCADIS Varic to be ready for operation again.
It is possible to switch on again in approx. 5 seconds after you used the switch
off button (even during shut down).
Switching off (simple shut down) can be used up to 15 times in a row. After the
tenth shut off using the OFF button, a dialog opens in which you can select sim-
ple or complete shut down. After the 15th switch off using the OFF button,
a complete shut down automatically occurs with appropriate notification.
If you want to ensure a very short power-up time of your ARCADIS Varic system,
we recommend that you initiate a complete shutdown yourself before the maxi-
mum number of 15 simple shutdown procedures is reached. You can do this dur-
ing a break between examinations, for example, and then switch the
ARCADIS Varic on again in time before the next examination.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...