Operator Manual
22
of 28
SPR2-310.620.30.02.02
Safety
The values are valid for characteristic HC2 high, 68 kV, 1.3 mA, C-arm horizontal,
downward beam direction without additional Cu filter, with scattered radiation
grid.
The values are valid for continuous fluoroscopy 110 kV, 2.9 mA, C-arm horizontal,
downward beam direction without additional Cu filter, with scattered radiation
grid.
Radiation interruption for all operating modes
0
The hand and footswitches are designed as pushbutton switches. Radiation is
interrupted immediately when releasing the corresponding operating element or,
in other operating modes, after the completion of the stored image.
Height above the floor
[cm]
Operating area A/
Measurement B
I.I. 23 cm/9"
110
< 1.4 mGy/h
Height above the floor
[cm]
Operating area B/
Measurement D
I.I. 23 cm/9"
140 to 200
< 41 µGy/h
Height above the floor
[cm]
Operating area A/
Measurement A
I.I. 23 cm/9"
110
< 9 µGy/h
Height above the floor
[cm]
Operating area B/
Measurement C
I.I. 23 cm/9"
140 to 200
< 300 µGy/h
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...