Operator Manual
24
of 24
SPR2-310.620.30.02.02
Options
Releasing the exposure
0
◆
Press the release button on the hand switch to release the direct exposure.
❏
The radiation indicators light up during the exposure. The radiation indicators
light up a little longer for very short exposure times so that radiation is clearly
indicated.
❏
During the exposure or after ending the exposure, an acoustic warning signal
sounds (can be configured).
Removing the cassette
0
◆
Take the cassette out of the cassette holder.
– After removing the cassette, the exposure data remains until another oper-
ating mode is selected.
Removing the cassette holder
0
◆
Hold the spring.
◆
Gently push the cassette holder downward, lifting it over the front ring to
remove it.
Switching the operating mode
0
You must remove the cassette before changing over to another operating mode.
◆
Select the desired operating mode.
Or
◆
Release fluoroscopy without selecting an operating mode.
– If no operating mode is selected, the system automatically reverts to the
operating mode selected prior to the cassette exposure.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...