Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02
5
of 34
(6) IOD 550 W
SIREMATIC IOD
Iodine-contrast optimized
standard mode max. 8.3 mA
in Power Mode 1000 W max. 15.2 mA
SIREMATIC curves
Fluoroscopy/Pulsed Fluoroscopy
S-1
Antiisowatt curve with max. 3 mA
S-2
Antiisowatt curve with max. 5 mA
LD
Low dose curve with high kV values: i.e. low radia-
tion exposure for the patient, especially well-suited
for pediatrics.
The characteristically high kV values result in lower
image contrast.
HC -1
High-contrast curve with max. 4.7 mA
HC -2
High-contrast curve for applications requiring higher
mA values (max. 7.8 mA) e.g. spinal column, hip,
skull
IOD
IODINE curve
For special applications using iodine contrast agent