Operator Manual
48
of 78
SPR2-310.620.30.02.02
Image Processing
Drawing a distance
line
◆
Move the mouse cursor into the image.
– The cursor changes shape.
◆
Click on the image to set the starting point and drag a line to the vertex by
using the mouse.
– As soon as you move the mouse, the line is displayed.
◆
Release the mouse button to finish the line.
– The distance line is displayed.
– The
Calibration
dialog box is displayed.
◆
Enter the distance (in cm).
◆
Confirm with
OK
.
– Distance calibration is performed and the calibration factor is displayed in
the image.
– Distances already drawn in the image are updated.
Canceling calibration
◆
Click
Cancel
.
– The previous calibration factor is maintained.
Deleting the distance
line
◆
Select
Edit > Remove Calibration
.
– The existing distance line is deleted.
– Instead of the current length, "
??
" is displayed in all measurement graphics.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...