Examination
ARCADIS Varic
SPR2-310.620.30.02.02
9
of 52
Radiation data (Extended display mode only)
0
The different dose levels of the selected exam set are displayed in a progress bar.
Here, the entire fluoroscopic time since the start of the examination of a patient
is displayed.
The cumulative area dose product for the current patient is displayed, if the
optional dose measurement chamber is installed.
As an alternative:
Display of air kerma values. The cumulated air kerma value is
identified by a preceding dot.
The reference location for determining the air kerma strength with the optional
dose measurement chamber is 30 cm in front of the image intensifier input.
The reason for this convention is that in typical applications the object to be ex-
amined is located approximately 30 cm in front of the I.I.
Depending on country-specific regulations, you can have Siemens Service
change the display so that it indicates the air kerma and cumulated air kerma
value instead of the area dose product.
Summary of Contents for ARCADIS Varic
Page 2: ......
Page 3: ...www siemens com healthcare ARCADIS Varic Operator Manual ...
Page 6: ......
Page 8: ......
Page 12: ...Operator Manual 2 of 8 SPR2 310 620 30 02 02 Introduction ...
Page 90: ...Operator Manual 44 of 44 SPR2 310 620 30 02 02 System Description ...
Page 150: ...Operator Manual 60 of 60 SPR2 310 620 30 02 02 Patient data ...
Page 284: ...Operator Manual 4 of 98 SPR2 310 620 30 02 02 Documentation ...
Page 380: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Configuration ...
Page 436: ...Operator Manual 34 of 34 SPR2 310 620 30 02 02 Technical Data ...
Page 438: ...Operator Manual 2 of 16 SPR2 310 620 30 02 02 Maintenance ...
Page 454: ...Operator Manual 2 of 24 SPR2 310 620 30 02 02 Options ...