Graphically representing power versus time (Scope mode)
R&S NRP
1144.1400.12 2.18
E-2
Graphically representing power versus time (Scope mode)
In the
Scope
mode, power versus time can be represented graphically as with an oscilloscope. If you
wish to carry out time-dependent measurements (in the Burst or Timeslot mode) on modulated signals,
displaying the signal in the Scope mode is always initially recommended since stable triggering by the
R&S NRP is most clearly recognized in this mode.
1
Setting the Scope mode.
¾
Connect a power sensor of the R&S NRP-Z1x or
R&S NRP-Z2x family to connector A of the R&S NRP
and apply an amplitude-modulated or pulsed signal
having a modulation or pulse frequency of approx.
1
kHz. The maximum level should be between
–10 dBm and +10 dBm.
({PRE}SET)
*
*
¾
Press
the
({PRE}SET)
key twice.
A measurement window is now available.
–
( )
*
↔
*
¾
Change to the
Sensor
menu and select
Mode
.
Mode …
–
( )
*
… T'slot T'gate Scope
–
( )
*
↔
*
¾
Switch to the
Scope
mode.
The associated parameters will appear in the lower part
of the dialog box. You can leave the parameters un-
changed for this exercise.
For a detailed description, see Chapter 4, Displaying data
in the Scope mode.
(¶ MENU)
*
¾
Close
the
Mode
dialog box.
–
( )
*
↔
*
¾
Change to the
Windows
menu and press
Expand
.