System settings (System menu)
R&S NRP
1144.1400.12 4.56
E-2
Remote control via USB
Remote
dialog box:
GPIB
USB
–
( )
ä
↔
ä
Fig. 4-52
Remote – Dialog, GPIB
Remote control of the R&S NRP via the USB is according to the USBTMC standard. For this purpose, a
VISA library that supports remote control via the USB is required by the remote-control computer. The
VISA library is equipped with the necessary USB device driver. The R&S NRP has a fixed, unchange-
able address comprising a manufacturer number (0x0AAD for Rohde & Schwarz), an instrument num-
ber (2 for the R&S NRP) and the serial number of the R&S NRP. The serial number can be displayed in
the System Info dialog box (
Information about the R&S NRP, page 4.63) and must be completely, as
displayed, included in the address. The VISA address string suitable for the device is displayed in the
dialog box.
Configuring the Ethernet
Remote control of the R&S NRP via Ethernet is according to the VXI-11 standard. Prerequisite for this
is the installation of the R&S NRP-B5 option. Due to the complexity of the protocol, the use of a VISA
library is also urgently recommended here. The VISA library that is used must support VXI-11.
The R&S NRP must have a unique IP address in the Ethernet. This address can be assigned either
manually or automatically using the DHCP protocol.
Remote
dialog box:
VXI11
Eth
–
( )
ä
↔
ä
Fig. 4-53
Remote – Dialog, GPIB
IP Address
–
( )
ä
ä
Opens the dialog box for the configuration of the IP address.
Remote : USB
Remote : VXI 11