Remote Control - Commands
R&S NRP
1144.1400.12 6.54
E-3
[SENSe[1..4]]:CORRection:OFFSet[?] <float_value>
With this command a fixed offset value can be defined for multiplying (logarithmically adding) the
measured value of a sensor. If the parameters are specified without a unit, the unit set by
UNIT:POW:RAT
will be used.
Unit:
DB | PCT
Default unit:
DB
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:CORRection:OFFSet[?]:STATe ON | OFF
This command switches the offset correction on or off.
*RST value
:
depending on sensor
[SENSe[1..4]]:CORRection:DCYCle[:INPut][:MAGNitude][?]
<float_value>
This command informs the R&S NRP about the duty cycle of the power to be measured. The duty cycle
correction is only applied in the ContAv mode where measurements are performed continuously without
taking the timing of the signal into account.
Unit:
PCT
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:CORRection:DCYCle:STATe[?] ON | OFF
This command switches duty cycle correction for the measured-value on or off.
*RST value
:
depending on sensor
[SENSe[1..4]]:CORRection:FDOFfset[:INPut][:MAGNitude]?
<float_value>
This command is a query only. The frequency-dependent measured-value correction is set by selecting
a table with the
[SENSe[1..4]]:CORRection:FDOTable[:SELect]
and
[SENSe[1..4]]:CORRection:FDOTable:STATe ON
commands. The respective frequency-
dependent correction factor is determined by means of the active tables (if necessary by interpolation)
and can be queried with
[SENSe[1..4]]:CORRection:FDOFfset[:INPut][:MAGNitude]?
(frequency-dependent offset).
Unit:
DB
*RST value
:
none