Remote Control - Commands
R&S NRP
1144.1400.12 6.52
E-3
[SENSe[1..4]]:AVERage:COUNt:AUTO:MTIMe[?] <float_value>
If the R&S NRP has to determine the filter length automatically, the filter length can become large and
thus also the measurement time.
An upper time limit can be set via
SENS[1..4]:AVER:COUN:AUTO:MTIME
(maximum time). It should
never be exceeded. Undesired long measurement times can thus be prevented (in the
SENS:AVER:COUN:AUTO:TYPE NSR
mode) if the automatic filter length switchover is on.
Unit:
S
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:COUNt:AUTO:RESolution[?] 1 | 2 | 3 | 4
Defines the number of significant places for linear units and the number of decimal places for
logarithmic units which should be free of noise in the measurement result. This setting is also performed
by the
DISP[1..4]:RES
command (
Æ
page
6.35
) that tries to set the sensors involved in the
measurement results accordingly. (For further details, see Chapter 4.)
The setting is only taken into account, if
SENS:AVER:COUN:AUTO:TYPE
is set to
RES
and
SENS:AVER:COUN:AUTO
is
ON.
However,
SENS[1..4]:AVER:COUN:AUTO:RES
does not affect the DISPlay command. The
parameters of the two commands are different but have the same meaning.
SENS[1..4]:AVER:COUN:AUTO:RES
1 2 3 4
DISP[1..4]:RES
1 0.1
0.01
0.001
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:COUNt:AUTO:SLOT[?] <int_value>
Sets a timeslot whose measured value is used to automatically determine the filter length.
Unit:
-
Value range:
depending on sensor
*RST value
:
depending on sensor