R&S NRP
Remote Control - Commands
1144.1400.12 6.39
E-3
MEMory
The MEMory system is used for storing device states (
:STATe
) and correction tables (
:TABLe
).
:STATe
To allow a rapid and convenient changeover between device setups for different applications,
up to 20 setups can be stored in the device. The setup with the number 0 is the default setup
made in the factory. This setup can neither be changed nor cleared.
:TABLe
Tables can be stored for a frequency-dependent correction of measured values. Up to 10
tables with two columns and up to 80 entries (lines) can be created. Thus, not only the
sensors but also the frequency response of various test assemblies can be compensated for
during the measurement.
The tables are selected with
[SENSe[1..4]]:CORRection:FDOT[:SELect]
and then
activated and deactivated with
[SENSe[1..4]]:CORRection:FDOT[?]:STATe
.
Names can be assigned to the tables. Some commands which do not specify a table name
affect the table currently selected with
MEM:TABL:SEL <name>
.
Note:
Stored data is not affected by an
*RST
command.
Table 6-7: Commands of the MEMory system
Command
Parameter
Unit
Remark
MEMory
:CATalog
[:ALL]?
-
-
Query only
:STATe?
-
-
Query only
:TABLe?
-
-
Query only
:CLEar
[:NAME]
<name>
-
No query
:TABLe
-
-
No query
:FREE
[:ALL]?
-
-
Query only
:STATe?
-
-
Query only
:TABLe?
-
-
Query only
:NSTates?
-
-
Query only
:STATe
:CATalog?
-
-
Query only
:DEFine|:MAP[?]
<string>, 0..19
-
:TABLe
:FREQuency[?]
<NRf>{,<NRf>}
HZ
:POINts?
-
-
Query only
:GAIN
[:MAGNitude][?]
<NRf>{,<NRf>}
DB|PCT
:POINts?
-
-
Query only
:MAP[?]
<string>, 0..9
-
:MOVE
<string>,<string>
-
No query
:SELect[?]
<string>
-