Remote Control - Commands
R&S NRP
1144.1400.12 6.102
E-3
SYSTem:STANdard:PRESet <string>
The R&S NRP can be easily configured for measuring standardized signals (e.g. mobile radio
standards). For this purpose, the standards listed in the Annex (
Æ
Standard Configurations
) are stored
in the R&S NRP. The
SYST:STAN:PRES
command sets the listed parameters and, if desired, also
configures front-panel display windows (if
SYST:STAN:PWS ON
).
*RST
value:
none
SYSTem:STANdard:PWSettings ON | OFF
If
SYST:STAN:PRES <string>
is used,
SYST:STAN:PWS
(Preserve Window Settings) determines
whether the window system is also configured with the calculate blocks (
OFF
) or whether exclusively
sensor parameters are set (
ON
).
*RST
value:
OFF
SYSTem:STANdard:TRIGger:SOURce INT | EXT
If
SYST:STAN:PRES <string>
is used, this command determines whether the sensors are set to
internal (
INT
) or external (
EXT
) triggering.
*RST
value:
INT
SYSTem:TIME[?] <hour>,<minute>,<second>
Parameter
Value range
<hour>
0 ... 24
<minute>
0 ... 59
<second>
0 ... 59
Sets the time. Since the R&S NRP has no battery-backed clock, the time has to be reset if the clock
was not set after the last power-on of the R&S
NRP. (
Æ
SYSTem:DATE[?]
<year>,<month>,<day>
).
SYSTem:VERSion?
Returns a string that contains the current version number of the SCPI standard.