R&S NRP
Remote Control - Commands
1144.1400.12 6.11
E-3
High-Level Measurement Commands (
CONFigure
,
MEASure
,
READ
and
FETCh
)
The high-level measurement commands
CONFigure
,
MEASure
,
READ
and
FETCh
combine several
setting commands and thus simplify programming of the R&S NRP. The most convenient command is
MEASure
. This command configures sensors and calculate blocks, initiates a measurement and
provides measurement results. The other commands perform only portions of these functions and thus
allow the user to make specific modifications between the individual steps.
Strictly speaking, the
CONFigure
command is not a measurement command because it does not yield
data and only configures a measurement. It differs from the other three high-level measurement
commands as follows: a question mark at the end of the command provides information on current
settings. With other commands, the question mark is compulsory and shows that the measured data is
queried.
The commands have different functions:
CONFigure:...
The calculate blocks and the associated sensors are configured by means of the
parameter list. A measurement is not started. A question mark at the end of the
command provides information on the parameters transferred by the last
CONFigure
or
MEASure
command, respectively. Since the device settings can be
changed after a
CONFigure
, the query does not return the current device setup
1
.
CONFigure
is performed internally by SENSe and CALCulate commands. One or
several measurement channels are configured as follows:
Measurement mode:
ContAv
Burst
Timeslot
Scope
INIT:CONT OFF
OFF
OFF
OFF
TRIG:SOUR IMM
INT
EXT
IMM
TRIG:COUN 1
1
1
1
TRIG:DEL:AUTO ON
ON
ON
ON
SENS:AVER:COUN:AUTO ON
ON
ON
-
SENS:AVER:COUN:AUTO:TYPE
ON ON ON -
SENS:AVER:COUN:AUTO:SLOT
- - 1 -
SENS:AVER:STAT ON
ON
ON
ON
UNIT:POW
UNIT:RAT
Is set as a function of the selected calculation
function.
1 See also Annex "
Compatibility Information
".