R&S NRP
Remote Control - Commands
1144.1400.12 6.67
E-3
SERVice[1..4]:RCOunt[?] <NR1>
SERVice[1..4]:RESult[?] <NR1>
These commands allow simulated measured values to be specified for test purposes.
SERV:RCO
specifies the number of
SERV:RES
pairs to follow. The first
SERV:RES
command specifies the
measured value in watts, the second
SERV:RES
command determines how often the measured value
should be obtained. The measurement sequences defined in this way are repeated until the sensor is
disconnected from the R&S NRP and reconnected.
Example:
The command sequence
SERV2:RCO 2
SERV2:RES 3
SERV2:RES 2
SERV2:RES 4
SERV2:RES 1
makes sensor 2 return the measured values 3 W, 3 W, 4 W (, 3 W, 3 W, 4W, ....) from now on.
*RST
value:
none. The simulation mode is not exited in the case of
*RST
.
SERVice[1..4]:UNLock[?] 1234
Before other SERVice system commands are accepted, they must be enabled with
SERV:UNL 1234
.