R&S NRP
Remote Control - Commands
1144.1400.12 6.23
E-3
CALCulate[1..8]:LIMit:BEEP ON | OFF
Switches the acoustic warning for limit violations on or off.
*RST
value:
OFF
CALCulate[1..8]:LIMit:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
CALCulate[1..8]:LIMit:CLEar:AUTO ON | OFF | ONCE
Automatically resets the limit monitoring state and the internal counter for limit violations if any of the
following events occurs:
•
INITiate[:IMMediate]
is executed
•
INITiate:CONTinuous ON
is executed
•
A
MEASure?
or
MEASure:...?
query is executed.
•
A
READ?
or
READ:...?
query is executed.
If the automatic function is set to
ONCE
, the counter is reset only when the next of these events occurs.
*RST
value:
OFF
CALCulate[1..8]:LIMit:FAIL?
Queries whether upper or lower limits have been exceeded. The status is reset by the following events
(0 = no limit violation):
•
The device is switched on
•
A reset is performed (
*RST
)
•
The
CALCulate[1..8]:LIMit:CLEar
command is sent.