R&S NRP
Remote Control - Commands
1144.1400.12 6.93
E-3
SYSTem
With the aid of the SYSTem system, administrative device settings can be made and queried. This
includes:
•
list of error messages
•
IEC/IEEE-bus
address
•
resetting the R&S NRP
•
setting the date and time
•
audible
signals
•
version numbers of hardware and software
•
setting the system speed
Table 6-31: Commands of the SENSe system
Command
Parameter
Unit
Remark
SYSTem
Query only
:BEEPer
[:IMMediate]
-
No query
:NOTify
:OVERload[?]
ON | OFF
-
:KEY[?]
ON | OFF
-
:STOP
-
-
No query
:TIME
<NRf>
S
No query
:COMMunicate
:GPIB
[:SELF]:ADDRess
0..30
No query
:DATE[?]
<year>,<month>,<day>
:INFO?
[<string >]
Query only
:ERRor?
-
Query only
:KEY
<NR1> | character_value>
No query
:PRESet
-
No query
:SENSor[1..4]
:INFO?
[<string >]
Query only
:RESet
-
No query
:TEST?
-
-
Query only
:SPEed
NORMal | FAST
No query
:STANdard
:CATalog?
-
-
No query
:PRESet
<string>
-
No query
:PWSettings[?]
ON|OFF
-
:TRIGger:SOURce[?]
INT|EXT
-
:TIME[?]
<hour>,<minute>,<second>
:VERSion?
-
Query only
SYSTem:BEEPer[:IMMediate]
Generates an audible signal via the built-in loudspeaker. The duration of the signal is set with
SYST:BEEP:TIME
. Frequency and volume cannot be changed. The signal is only output if it is not
switched off with
SYS:BEEP:STAT OFF
.
SYSTem:BEEPer:NOTify:OVERload[?] ON | OFF
Determines if the R&S NRP shall emit an acoustic signal when the power at a sensor exceeds the
power limit.
*RST
-Wert:
OFF