TRIGger[1..4|:ALL][:IMMediate]
Performs triggering and ensures that the sensor directly changes from the WAIT_FOR_TRG state to
the MEASURING state irrespective of the selected trigger source (
TRIG:SOUR
). A trigger delay set with
TRIG:DEL
is ignored but not the automatic delay determined when
TRIG:DEL:AUTO:ON
is set.
When the trigger source is HOLD, a measurement can only be started with TRIG.
Error messages:
-211 "
Trigger ignored
": The sensor is not in the WAIT_FOR_TRG state.
TRIGger[1..4|:ALL]:LEVel[?] <float_value>
Determines the power a trigger signal must exceed (
TRIG:SLOP POS
) or fall short of (
TRIG:SLOP
NEG
) before a trigger event is detected. This setting is only used for the trigger signal source
(
TRIG:SOUR
)
INTernal
.
Unit:
DBM | W
Default unit:
DBM
Value range:
depending on sensor
*RST value
:
depending on sensor
TRIGger[1..4|:ALL]:SLOPe[?] POSitive | NEGative
This command determines whether the rising (
POSitive
) or the falling (
NEGative
) edge of the signal
is used for triggering.
*RST value
: