R&S NRP
Remote Control - Commands
1144.1400.12 6.53
E-3
[SENSe[1..4]]:AVERage:COUNt:AUTO:NSRatio[?] <float_value>
Indicates the maximum noise ratio in the measurement result.Achieving the same result by filter length
setting is attempted only if
SENS[1..4]:AVER:COUN:AUTO ON
and
SENS[1..4]:AVER:COUN:AUTO:TYPE NSR
are set.
Unit:
DB | PCT
Default unit:
DB
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:COUNt:AUTO:TYPE[?] RESolution | NSR
Selects a method by which the automatic filter length switchover can operate:
SENS[1..4]:AVER:COUN:AUTO:TYPE
Setting taken into account
RESolution SENS[1..4]:AVER:COUN:AUTO:RES
NSR SENS[1..4]:AVER:COUN:AUTO:NSR
*RST value
:
depending on sensor
[SENSe[1..4]]:AVERage:RESet
Initializes the digital filter by deleting the stored measured values.
*RST
value:
none
[SENSe[1..4]]:AVERage:TCONtrol[?] MOVing | REPeat
As soon as a new single value is determined, the filter window is advanced by one value so that the
new value is taken into account by the filter and the oldest value is forgotten.
[SENSe[1..4]]:AVERage:TCONtrol
(terminal control) then determines whether a new result is
calculated immediately after a new measured value is available (
MOVing
) or only after an entire range of
new values is available for the filter (
REPeat
).
*RST
-Wert:
(PRES)
and
SYST:PRES
:
MOV
*RST
:
REP
Note:
In the
MOVing
mode, data from a point of time before the start of the
current measurement can influence the measurement result if the data is
still in the filter.