R&S NRP
Installation in a 19" rack
1144.1400.12 1.7
E-2
Installation in a 19" rack
Caution
If the meter is rack-mounted ensure there is a free flow of air through the holes in the
side panels.
The R&S NRP can be installed in 19" racks using a variety of rack adapters (see data sheet for order
Nos.). The installation instructions come with the adapter.
AC supply voltage
The R&S NRP can be operated from AC lines with a voltage range of 100 V to 240 V and a frequency
range of 50 Hz to 60 Hz. Note that a restricted voltage range (100 V to 120 V) applies to 400 Hz net-
works. The AC supply connector is at the rear of the meter. The meter sets itself automatically to the
applied voltage if it is in range.
AC supply fuses
The R&S NRP has two fuses as indicated by the type plate. The fuses are accommodated in a fuse
holder in the AC supply connector. The holder can be pulled out to insert fuses. The power supply has
its own fuse.
Caution
Only the service department may replace the internal fuse.
AC supply connector
Fuse holder
AC supply connector at the rear of the meter
EMC
To prevent EMI, the meter must always be installed to meet the relevant EMC standards. Never operate
the instrument with its enclosure removed. Only use shielded signal and control cables that meet the
relevant EMC standards.