R&S NRP
Supplement
1144.1400.12-04- Supplement
3 E-
2
(¶ MENU)
ä
The calibration cycle is performed. First of all, the battery is fully
charged, then fully discharged and finally fully charged again. This
process may take between 11 and 20 hours depending on the charge
status of the battery at the beginning of the calibration cycle. The
higher the charge of the battery at the beginning, the faster the
calibration cycle is performed. Calibration itself takes place at the end
of the discharging process.
During the calibration cycle, or at least until the end of the discharging
phase, the R&S NRP should remain connected to the AC supply and
should always be on. Otherwise the calibration counter is not calibra-
ted, which means that the warning will be displayed again when the
R&S NRP is switched on the next time.
(ESC/LOCAL)
ä
The calibration cycle is not performed. The warning will be displayed
again when the R&S NRP is switched on the next time.