R&S NRP
Remote Control - Commands
1144.1400.12 6.35
E-3
DISPlay[:WINDow][1..4]:RESolution[?] 1 | 0.1 | 0.01 | 0.001
Specifies the number of significant places of the mantissa when the measurement result is displayed on
a linear scale. The R&S NRP tries to adjust the averaging filter range so that the display noise
decreases with increasing resolution. The values of the transfer parameter have the following meaning:
Parameter
Linear scale of measurement result
Logarithmic scale of measurement result
1
1 significant place
No decimal place
0.1
2 significant places
1 decimal place
0.01
3 significant places
2 decimal places
0.001
4 significant places
3 decimal places
Note:
Automatic filter ranging must be switched on
(
SENSe[1..4]:AVERage:COUNt:AUTO ON
). DISP[1..8]:RES has an
effect on the setting
SENS[1..4]:AVER:COUN:AUTO:RES
of the sensors
used in the measurement function (
CALC[1..8]:MATH
). However, the
resolution setting of the sensors does not affect the resolution of values in
the display.
This command corresponds to the
<resolution>
option of the high-level measurement commands
CONFigure
,
FETCH
,
READ
and
MEASure.
*RST value
:
0.01
Error messages:
-127 "
Invalid numeric data
": An invalid number was specified for the resolution.
Example
:
DISP:WIND1:RES 0.1
With a measured value of 123.456789 W, the first two places are free of noise and thus 120 W is
displayed.
DISPlay[:WINDow][1..4][:STATe][?] ON | OFF
Opens or closes a window. When a window is closed the other windows are automatically expanded so
that the free space is utilized. Likewise, open windows become smaller when a new one is opened.
*RST value
:
ON
(Window 1)
OFF
(Window 2,3,4)