Displaying measurement results (Windows menu)
R&S NRP
1144.1400.12 4.30
E-2
Configuring the result display
Each of the display modes has special configuration parameters.
In the digital display, the resolution is adjustable (
Resolution, page
4.30).
In the hybrid digital/analog mode, you can adjust the resolution of the
digital display (
Resolution, page 4.30) and specify the scaling for the
analog scale (
Scale, page 4.31).
In the Graph display mode, there are no other parameters in the menu.
The many possible settings in this display mode are made directly in
the window, which for this purpose has to be expanded (
Expanding
windows
(Expand)
, p. 4.29).
Fig. 4-26
Windows menu, display-specific softkeys
Setting the resolution
The resolution of the digital measurement result can be set in 4 stages, designated as 1 dB, 0.1 dB,
0.01 dB and 0.001 dB. If dB, dBm or dB
µ
V is selected as the unit, the number of places after the deci-
mal point is set directly. In the linear mode (W,
∆
%, 1), the number of digits after the decimal point de-
pends on the resolution, a digit being added or removed for each resolution stage that is added or re-
moved.
The selected resolution also influences the action of the autofilter in the
The
Normal
mode (see page
4.19). The greater the selected resolution, the longer the averaging filter to reflect the number of digits.
Windows menu
:
Resolution (dB)
–
( )
1
0.1 0.01 0.001
ä
↔
ä
Sets the result resolution you want.
Remote Control
DISPlay[:WINDow][1..4]:RESolution[?] 1 | 0.1 | 0.01 | 0.001
Windows