Remote Control - Commands
R&S NRP
1144.1400.12 6.64
E-3
[SENSe[1..4]]:TRACe:REALtime[?] ON | OFF
In the default state (
OFF
), each measurement from the sensor is averaged. With
[SENSe[1..4]]:TRACe::REALtime ON
,
this behaviour can be switched off, which may increase
the measurement speed. This ensures that the measured values of an individual measurement are
immediately delivered. (See also Chapter 4.)
*RST value
:
depending on sensor
[SENSe[1..4]]:TRACe:TIME[?] <float_value>
Sets the time to be covered by the Scope sequence.
Unit:
S
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:TIMing:EXCLude:STARt[?] <float_value>
Sets a time that is to be excluded at the beginning of the integration (
Æ
Fig. 6-9
).
Unit:
S
Value range:
depending on sensor
*RST value
:
depending on sensor
[SENSe[1..4]]:TIMing:EXCLude:STOP[?] <float_value>
Sets a time that is to be excluded at the beginning of the integration (
Æ
Fig. 6-9
).
Unit:
S
Value range:
depending on sensor
*RST value
:
depending on sensor