Configuring measurements (Measurement menu)
R&S NRP
1144.1400.12 4.32
E-2
Configuring measurements
Measurement windows
For every window opened via the
Windows
menu, you can configure a measurement in the
Measure-
ment
menu. The measurement channels (A, B, C and D), how the measured values are processed
(ratio, difference), the unit for the result, additional information (max, min), etc are also defined. This
procedure does not change the sensor operating modes that have been set in the Mode menu. All key
parameters are displayed along with the result in the appropriate window.
In the manual mode, four measurements can be configured, corresponding to the maximum number of
windows (1, 2, 3 and 4); in the remote control mode, additional measurements 5 to 8 can be configured
(not displayed on screen).
The Measurement menu
The Measurement menu contains the functions required to configure measurements.
Selecting a window, page 4.32.
Selecting a measurement function and the unit, page 4.33.
Relative measurements, page 4.39.
Setting the reference value manually, page 4.39.
Limit monitoring, page 4.40.
Resetting the extreme value display, page 4.38.
Fig. 4-29
Measurement menu
Selecting a window
You can select one of the windows (1 to 4) as and when you want. All
Windows
and
Measurement
menu
functions act on the selected window.
Measurement menu
:
Window –
( )
1
2 3 4
ä
↔
ä
Selects the window whose number is displayed in the menu. If the
window is open, this is also indicated by the window’s title bar being in
inverse video and by the “shadow” along the window’s right and bottom
edges.
Measurement
Window