Configuring measurements (Measurement menu)
R&S NRP
1144.1400.12 4.40
E-2
Limit monitoring
An upper and a lower limit for each window can be set in the Limits… dialog box. If either limit is vio-
lated, a warning is issued. The warning may be one of the following:
•
The
LIMIT FAIL
message is displayed on the screen.
•
A warning tone is output.
•
A signal is issued on analog output OUT1 (
Setting the analog outputs on page 4.58).
Measurement menu
:
Limits… –
( )
ä
Fig. 4-34
Limits dialog box
Upper
[ß ]
–
( )
ä
Turns monitoring of the upper limit on/off.
Remote Control
CALCulate[1..8]:LIMit:UPPer:STATe[?] ON | OFF
[ .... ]
Upper
–
( )
ä
Upper
limit.
Remote Control
CALCulate[1..8]:LIMit:UPPer[:DATA][?] <float_value>
Lower
[ß ]
–
( )
ä
Turns monitoring of the lower limit on/off.
Remote Control
CALCulate[1..8]:LIMit:LOWer:STATe[?] ON | OFF
[ .... ]
Lower
–
( )
ä
Lower
limit
Remote Control
CALCulate[1..8]:LIMit:LOWer[:DATA][?] <float_value>
Warning Beep
–
( )
ä
↔
ä
For activating the acoustic alarm.
Remote Control
CALCulate[1..8]:LIMit:BEEP[?] ON | OFF
Measurement
Limits